Repair of electronic floor scales CAS DL
CAS DL floor scale repair in Moscow and the region — official dealer

Repair of CAS DL and CAS DL-N floor scales in Moscow and the Moscow region. We take 60, 100, 150 and 200 kg models: the head will not switch on, the display shows Err, Init, OFL or UNStA, the weight drifts, buttons fail, the glass is broken, the pole is bent. Repair of any complexity, spare parts in stock, calibration, verification, on-site visit, collection and return delivery. Payment by bank transfer.

Official CAS dealer and authorized service center. Here you can not only repair CAS DL scales, verify CAS DL scales and buy CAS DL spare parts, but also buy new scales with verification and warranty.

CAS DL floor scale: stainless platform, pole and LCD head
CAS DL: 370×500 mm platform, pole, LCD head. The current version has 4 buttons.

In short. CAS DL is a floor scale with a stainless-steel platform and an LCD indicator on a pole. Range: DL-60, DL-100, DL-150, DL-200, including the DL-N version. Current heads have 4 buttons; those scales are still supplied and the main board can be replaced. Scales with a 6-button board are discontinued and no longer supplied: the board can be repaired only if the processor is alive — processors are not available. Remont Vesov repairs them in the workshop and on site in Moscow and the region. Spare parts: spare parts for CAS DL-N floor scales.

CAS DL head with 4 buttons, current version

Current head, 4 buttons. The board is replaced, scales are still supplied.

CAS DL head with 6 buttons, discontinued

Old head, 6 buttons. Discontinued: the board is repaired only if the processor is alive.

Which CAS DL scales we repair

CAS DL floor scales are used in shops, warehouses, workshops and goods-in. Platform 370×500 mm, mains adapter or six C (R14) batteries, RS-232, tare, counting and checkweighing. Rated temperature −10…+40 °C.

Check capacity and division on the nameplate: some years differ. Typical range:

Model Capacity Division, typical Where used
CAS DL-60 / DL-60N 60 kg 20 g counter, small wholesale, packing
CAS DL-100 / DL-100N 100 kg 50 g warehouse, production, trade
CAS DL-150 / DL-150N 150 kg 50 g bags, cartons, meat and produce
CAS DL-200 / DL-200N 200 kg 100 g goods-in, heavy items

Division is checked on the head: some DL-150 units use 100 g. Capacity sets the calibration weight and the verification tolerance.

CAS DL-60 floor scale, 60 kg capacity

CAS DL-60, 60 kg, division usually 20 g.

CAS DL-100 floor scale, 100 kg capacity

CAS DL-100, 100 kg, division usually 50 g.

CAS DL-150 floor scale, 150 kg capacity

CAS DL-150, 150 kg, division usually 50 g.

CAS DL-200 floor scale, 200 kg capacity

CAS DL-200, 200 kg, division usually 100 g.

What the repair includes

  • Diagnosis of the head, load cell, pole cable, power supply and platform mechanics.
  • Repair of any complexity, from a button overlay and glass to the main and analogue boards.
  • Parts from stock — CAS DL-N spare parts, no wait on common items.
  • Zero and span calibration after repair, impact or overload.
  • Verification of trade and commercial scales — we do it ourselves, no need to send the scale elsewhere.
  • Engineer visit in Moscow and the Moscow region.
  • Collection for repair and delivery back to the site.
  • Payment by bank transfer for companies and sole traders.
  • Sale of new CAS DL scales with initial verification and warranty, as an official dealer.

Indicator error codes

On CAS DL the message is short: Err, Init, OFL, UNStA, BAT. Err has no number. Some clear on site — empty platform, level floor, live batteries. If the code stays, it is the load cell, the analogue path or a lost calibration.

CAS DL indicator showing Err
Err: initial zero is not set. Empty the platform and check the cover sits in the guides.
CAS DL indicator showing UNStA
UNStA: the signal is unstable, the scale will not lock the weight.

UNStA (some heads show UNSTA) means the reading will not settle. Check the floor, draughts, and whether the platform cover rubs the frame or debris sits under the cover. If the code stays on an empty level platform, we check the load cell, the pole cable, the analogue board and the stop bolts.

Code Meaning What to check What we do
Err Initial zero is not set at power-on Remove the load, check the cover sits in the guides and does not rub, press zero Calibration. If zero will not hold — load cell or analogue board
BAT Batteries flat Fit six fresh C cells or connect the adapter DC jack, adapter, battery contacts, compartment cover
OFL Platform overload Remove the load at once. Do not load DL-60 above 60 kg or DL-200 above 200 kg If the code is on an empty platform — cell shifted after an impact; replace or calibrate
UNStA Unstable signal Level the scale, stop draughts and contact with the frame, clear debris under the cover Load cell, analogue board, platform stop bolts
Init Analogue module fault Power off and on again. It will not clear by itself CAS DL-N analogue board, main board if needed

Without a code: weight looks right up to 40–60 kg and then jumps by kilograms — usually a stop bolt under the platform or a tired load cell. A segment stays lit, or the 1111…9999 test ends in Err — we check the indicator and the main board, not only the buttons.

Spare parts from stock

Items from spare parts for CAS DL-N floor scales. Catalogue prices are not fixed: they follow the dollar rate. Check the current price on the product page. A sharp move in the rate can change both the part price and the repair total. The 4-button board is replaced. Scales with a 6-button board are discontinued and no longer supplied: we do not fit a new board, we repair the original one, and only if the processor has not failed — processors are not available.

Electronics and display

Part When we replace it Price
Main board DL (4 buttons) Head is dead, random symbols, calibration will not store, current 4-key version see link
Main board DL (6 buttons) Old 6-button head. Scales discontinued and not supplied. Board is not replaced: we repair it if the processor is alive. No processors; a failed processor cannot be restored see link
Analogue board CAS DL-N Init, no response to load, Err with a good load cell see link
Indicator DL (22 legs) Dead segments, 22-pin indicator does not match the board see link
Indicator DL (26 legs) Same fault on a 26-pin indicator see link
Display glass DL-60 Cracked or fogged glass on a 60 kg head see link
Display glass DL-100 Broken glass on DL-100 see link
Display glass DL-150 Broken glass on DL-150. For DL-200 the glass is matched to the head see link

Power and keypad

Part When we replace it Price
CAS adapter Dead on mains, alive on batteries. Adapter from the DL section, same type as used on CAS SW see link
DC JACK socket Plug is loose, contact drops out, socket torn from the case see link
DL battery cover Cover lost or broken, batteries not held see link
Button CAS AD/DB/DL Key fallen in, Zero or Tare will not press see link
DL button cover One key overlay worn or peeled off see link
DL keypad overlay (6 buttons) Worn or torn membrane keypad on a six-button head see link

Housing, pole, fixings

Part When we replace it Price
DL pole bracket Pole hit by a trolley, head is loose see link
DL pole bracket cover Trim cover missing, fixing exposed see link
DL upper housing Top of the head broken, glass will not hold see link
DL rear housing Rear cover cracked, board not enclosed see link
Left indicator mount / right Head loose on the pole, mounting lug broken see link
DL metal nut Pole nut lost see link
DL nut holder plate Pole fixing will not hold, plate bent see link

The load cell is not in this list: it is matched to capacity and bridge resistance after measurement. The pole cable is checked for an open circuit before boards are replaced.

Calibration

We calibrate after a load cell, analogue or main board change, after overload, and when the scale is consistently wrong with sound mechanics. Zero and span are set with a weight close to capacity: up to 60 kg for DL-60, 100 kg for DL-100, 150 kg for DL-150, 200 kg for DL-200.

  • Platform empty, scale level, stops not jammed.
  • Zero first, then the reference load.
  • Check at several points and at the corners: the reading must not depend on which corner the load sits on.
  • On trade scales we verify after calibration. No need to send the scale to another organisation: we verify it ourselves.

The service entry for calibration is not published: on commercial scales the engineer does it, otherwise the tolerance is easy to lose.

Verification

If a CAS DL is used on the shop floor, at dispatch or in any account where the reading is used for payment, it needs verification. There is no need to send the scale elsewhere: we verify it ourselves, in the workshop and on site. After repair we:

  • clear the fault and fit the part;
  • calibrate to the model capacity;
  • check repeatability and corners;
  • verify the scale and issue the verification.

The verification period and the passport mark are on the certificate. An overdue scale is better kept out of trade until it is repaired and verified.

Visit, collection and delivery

Moscow and the Moscow region. Two options.

  1. Visit to the customer. The engineer checks the scale on site: level, platform, power, display code. Glass, buttons, adapter, socket and calibration are often closed on site. We bring the board and load cell if the fault is clear by phone.
  2. Collection and delivery. We collect the head or the whole scale, repair it in the workshop and bring it back. Better when the main board must be replaced, the socket soldered, or the shop floor is too busy to calibrate.

For a request we need the nameplate model (DL-60, DL-100, DL-150, DL-200 or DL-N), the display code and the address. Say whether the head has 4 or 6 buttons. A 4-button board is replaced. A 6-button scale is discontinued: we check whether the processor is alive.

Questions

How does DL differ from DL-N?

DL-N is the same floor series. For repair the button count and capacity matter more. The 4-button board is the current version and is replaced. Scales with a 6-button board are discontinued and not supplied. Glass is tied to the 60 / 100 / 150 model.

Do you repair 6-button scales?

Scales with a 6-button board are discontinued and no longer supplied. We do not fit a new board. Repair is only of the board itself, and only if the processor has not failed: processors are not available. A failed processor cannot be restored.

Do you verify the scales yourselves?

Yes. Calibration and verification are done by us, in the workshop and on site. There is no need to send the scale to another organisation.

The scale shows Err on an empty platform. Is that the batteries?

No. Err is the initial-zero fault, with no number. Batteries show BAT. Remove the load and check the cover. If it stays, bring the scale for calibration; often the load cell is the real cause.

Can you replace only the glass?

Yes, if the segments are alive. DL-60, DL-100 and DL-150 glass is in stock; links are in the table above.

Do parts have to be ordered in advance?

Common DL-N parts are in stock: boards, indicators, glass, pole, buttons, adapter and socket. Full list in the CAS DL-N spare parts catalogue.

How to contact us

Remont Vesov is an official CAS dealer and an authorized service center. Address, directions and bank details are on the remont-vesov.com home page. Phone +7 (999) 902-48-08, email zakaz@remont-vesov.com. Repair, verification and sale of CAS DL and DL-N floor scales in Moscow and the Moscow region: parts from stock, calibration, new scales with verification and warranty, visit, collection and delivery. Part and repair prices are on the catalogue links: they follow the dollar rate, and a sharp move can change the total.

CAS DL-60, DL-100, DL-150 and DL-200 are standard floor scales with a stainless platform. If the display shows Err, Init, OFL or UNStA, the head will not start from the mains or the pole is broken, we repair with a board, indicator, glass and fixings from stock. Scales with a 6-button board are discontinued: the board is repaired only if the processor is alive. After repair — calibration to capacity and verification. Part prices are on the links: they depend on the dollar rate.

Adapter for scales CAS SW

1,624.14р.

Adapter for scales CAS SW..

Adapter for scales CAS SW Adapter for scales CAS SW

Analogue board CAS DL-N

4,274.06р.

Analog board for CAS DL-N scales..

Analogue board CAS DL-N Analogue board CAS DL-N

Battery compartment cover DL for scales CAS

683.85р.

Battery compartment cover DL for scales CAS..

Battery compartment cover DL for scales CAS

Button cover DL for scales CAS

170.96р.

Button cover DL for scales CAS..

Button cover DL for scales CAS

Button for CAS AD/DB/DL scales

59.84р.

Button for CAS AD/DB/DL scales..

Button for CAS AD/DB/DL scales

Display glass DL-100 for scales CAS

427.41р.

Display glass DL-100 for scales CAS..

Display glass DL-100 for scales CAS

Display glass DL-150 for scales CAS

427.41р.

Display glass DL-150 for scales CAS..

Display glass DL-150 for scales CAS

Display glass DL-60 for scales CAS

427.41р.

Display glass DL-60 for scales CAS..

Display glass DL-60 for scales CAS

Indicator DL (22 legs) for scales CAS

1,453.18р.

Indicator DL (22 legs) for scales CAS..

Indicator DL (22 legs) for scales CAS

Indicator DL (26 legs) for scales CAS

1,025.77р.

Indicator DL (26 legs) for scales CAS..

Indicator DL (26 legs) for scales CAS

JACK DC CAS Scales

341.92р.

JACK DC 7610SBT0303B..

JACK DC CAS Scales

Keyboard overlay DL (6 buttons) for scales CAS

598.37р.

Keyboard overlay DL (6 buttons) for scales CAS..

Keyboard overlay DL (6 buttons) for scales CAS

Left indicator mount DL for scales CAS

170.96р.

Left indicator mount DL for scales CAS..

Left indicator mount DL for scales CAS

Load cell BC-150 for scales CAS

10,257.73р.

Load cell BC-150 for scales CAS..

Load cell BC-150 for scales CAS

Load cell BC-60 for scales CAS

10,257.73р.

Load cell BC-60 for scales CAS..

Load cell BC-60 for scales CAS
Showing 1 to 15 of 27 (2 Pages)